Battery analyzer

Battery analyzers are mainly based onElectrochemical principlesBy using methods such as cyclic charge/discharge, AC impedance spectrum (EIS), or constant-current intermittent titration (GITT), the voltage, current, temperature, and internal impedance parameters of the battery can be measured. This instrument can analyze the charge state (SOC), health status (SOH), capacity decay, internal resistance, and gas generation conditions in real time, and is commonly used in battery design, quality control, and life assessment.

10mA

The eight-channel battery analyzer can analyze small button and soft-pack batteries with current ranges from 0.02 mA to 10 mA and voltage up to 5 V. Each channel of the analyzer is equipped with an independent constant current source and constant voltage source, which can be programmed and controlled through computer software. This system can be widely applied in battery testing fields, such as electrode material research, battery performance testing, small-scale battery formation, performance grading, battery assembly testing, etc. The new notebook has been installed with the Windows 10 operating system and the latest version of controller software, and the analyzer has been calibrated and is ready for immediate use.

Product name/model8 Channel Battery Analyzer
Electricity requirementAC110~220V
Power Consumption120W
Charge & Discharge CurrentRange: 0.02 mA – 10mA Accuracy: ±(0.05% of reading + 0.1% of range)
Charge and discharge voltageRange: 0.5 – 5V programmable Accuracy: ±(0.05% of reading + 0.1% of range)
Data register conditionsTime interval: 1 – 999999s Voltage change: up to 5V Current change: up to 10mA
Max. measurement cycles65535 cycles 5 Nested Loops Maximum
ChannelsEight independent programmable channels Each channel can set different working modes and functions independently
Battery HoldersOne type of battery holder is included in the BST8-MA standard package: 8-Channel clips sets for coin cell. (Pic. 1) Optional battery holder: 8-Channel alligator clips sets for cylindrical cells. (Pic. 2) 8-Channel clip sets with bayonet connectors for pouch cell. (Pic. 3)
Programs & SoftwareThe software (Extreme Test) is included to set various working modes for measuring capacity and life cycle for all types of rechargeable batteries Working modes Include constant current discharge, constant current charge, constant voltage charge, constant resistance discharge, rest, etc Limited threshold conditions include voltage, current, time, capacity, negative voltage slope, etc With real-time monitoring windows and integrated graph/data windows, the testing process can be observed more directly and efficiently During the test, the software will provide instructions and warnings for assistance The calibration software can be used to calibrate the analyzer
Test reports and curves for analysisDifferent types of curves can be created by software based on user definition. (Voltage-time curve, current-time curve, capacity-voltage curve, loops times charge/discharge capacity curve, loops times charge/ discharge efficiency curve, etc) Data reports are created by the software. Users can easily compare the performance of the batteries tested in channels both visually and statistically
Protection and Auto-recoveryIf power failures occur during testing, the system will shut down all operating channels. Once power is recovered, the system will automatically reconnect with the PC, and the program can be resumed/continued manually.
Laptop ComputerBrand new laptop ( 4GB RAM 128GB SSD) with MS Window 10, and the latest version testing software is installed and calibrated for immediate use. Notice: It is suggested to use an external hard drive to keep an instant backup of the collected data.
ComplianceCE Certified NRTL certification is available upon request at an additional cost.
Application NotesNew buyers must buy the installed software on a laptop to avoid trouble during installation. If you want to run multiple battery analyzers on one laptop, you can simply buy multiple analyzers with one laptop, which are offered at a discounted price. Warning: Please choose NEVER for Screen and Sleep in the Power& Sleep setting of your computer, or it may lose connection from the battery analyzer. Multiple battery analyzers can be connected to 1 PC via an ethernet switch ($100.00). Since August 1, 2021, MTI has changed the software for better operation.

50mA

The eight-channel battery analyzer can analyze small button cells and soft gel batteries with current ranges from 0.1 mA to 50 mA and voltage up to 5 V.

Battery tester
*ModelUCET50
Current voltage5V 50mA
*General items
Input powerAC220V ± 10%, 50/60Hz
Input active power20W
AD/DAAD: 24 bit DA: 16 bit
Input impedance≧1GΩ
Noise<55dB
Protection levelIP20
Test sampling methodFour lines
Number of channels 8
※ Voltage
Range10mV~5V
Discharge voltage (min.)-5V
Accuracy±0.01% of FS
Stability0.02% of FS
Current
Export rangeRange 1: 0.1mA Range 2: 1mA Range 3: 10mA Range 4: 50mA
Minimum output current0.2 µA
Accuracy±0.01% of FS
Current outputRange 1: 0.1µA Range 1: 1µA Range 1: 10µA Range 1: 50µA
Stability0.02% of FS
* Power
Output power (max.)0.25W
Accuracy±0.02% of FS
Stability0.04% of FS
*Time 
Current response time≦1ms (10%-90% of FS)
Charging and discharging conversion time≦40ms (-90%-90% of FS)
Sampling frequency10 Hz (minimum time interval: 100 ms)
* Working mode
Charging and discharging modeConstant current charging and discharging, constant voltage charging and discharging, constant current and constant voltage charging and discharging, constant power charging and discharging, pulse charging and discharging, constant resistance charging and discharging, and follow-mode charging and discharging
Criteria as of nowVoltage, current, relative time, capacitance, energy, -△V customisable variables
DCIRCalculations are made at the self-pickup point
Pulse-wave charge-discharge modeConstant current mode, constant power mode
Minimum pulse width500ms
Number of pulse wavesA single pulse sequence supports 32 different pulses.
Pulse-wave switchingCharging continuous switching, discharging continuous switching
Pulse wave cutoff conditionVoltage, relative to time
*Circulation
Circulating test range1 to 65535 times
Single-loop sequence≦255
Nested cycle<10 layers
*Protection
Software protectionPower outage data protection, voltage upper and lower limit, current upper and lower limit, and custom variables…
Hardware protectionAnti-surge protection, over-voltage protection, over-current protection, over-temperature protection
*Communication requirements
CPU/RAM/Hard DrivePentium 4GHz dual core / 16G (or more) / 500GB or more
Communication protocolTCP/IP
Operating systemWIN7 or higher
DATABASEMySQL
Data export formatEXCEL,TXT,GRAPH,PDF,CSV
Other communication methods (optional)Support for CAN, RS485, BMS, DBC
*Environmental requirements
Working temperature range/humidity0℃~40℃ / ≤ 70% RH
Storage temperature range/humidity-10℃~50℃ ≦ 80% RH
Dimensions483x321x49 mm (19″ instrument rack)